Medical Policy
Policy Num: 01.001.018
Policy Name: Oscillatory Devices for the Treatment of Cystic Fibrosis and Other Respiratory Conditions
Policy ID: [01.001.018] [Ac / B / M + / P+] [1.01.15]
Last Review: September 09, 2024
Next Review: July 20, 2025
Related Policies: None
Population Reference No. | Populations | Interventions | Comparators | Outcomes |
1 | Individuals: · With cystic fibrosis | Interventions of interest are: · Oscillatory devices | Comparators of interest are: · Standard chest physical therapy | Relevant outcomes include:
|
2 | Individuals: · With bronchiectasis | Interventions of interest are: · Oscillatory devices | Comparators of interest are · Standard chest physical therapy | Relevant outcomes include:
|
3 | Individuals: · With chronic obstructive pulmonary disease | Interventions of interest are: · Oscillatory devices | Comparators of interest are: · Standard therapy | Relevant outcomes include:
|
4 | Individuals: · With respiratory conditions related to neuromuscular disorders | Interventions of interest are: · Oscillatory devices | Comparators of interest are: · Standard therapy | Relevant outcomes include:
|
Oscillatory devices are alternatives to the standard daily percussion and postural drainage method of airway clearance for patients with cystic fibrosis. There are several types of devices including high-frequency chest compression with an inflatable vest and oscillating positive expiratory pressure devices, such as the Flutter and Acapella devices. Respiratory therapists and other providers may also use oscillatory devices for other respiratory conditions such as diffuse bronchiectasis, chronic obstructive pulmonary disease (COPD), and respiratory conditions associated with neuromuscular disorders.
For individuals who have cystic fibrosis who receive oscillatory devices, the evidence includes randomized controlled trials (RCTs) and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. The RCTs reported mixed findings and limitations such as small sample sizes and large dropout rates. A systematic review identified 39 RCTs comparing oscillatory devices with other recognized airway clearance techniques; some were published only as abstracts. Reviewers could not pool findings due to heterogeneity in study designs and outcome measures and concluded that additional adequately powered RCTs with long-term follow-up would be needed to make conclusions about oscillatory devices for cystic fibrosis. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
For individuals who have bronchiectasis who receive oscillatory devices, the evidence includes RCTs and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. A 2015 systematic review identified 7 small RCTs on several types of oscillatory devices; only 1 reported the clinically important outcomes of exacerbations or hospitalizations. Only 3 RCTs reported on quality of life, and findings were mixed. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
For individuals who have chronic obstructive pulmonary disease (COPD) who receive oscillatory devices, the evidence includes RCTs and systematic reviews. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. Only a few controlled studies have evaluated oscillatory devices for the treatment of COPD, and they tend to have small sample sizes, short follow-up periods, and limitations in their analyses (eg, lack of intention-to-treat analysis and between-group comparisons). Moreover, the published studies reported mixed findings and did not consistently support the use of oscillatory devices in this population. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
For individuals who have respiratory conditions related to neuromuscular disorders who receive oscillatory devices, the evidence includes 2 RCTs and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. One of the RCTs was not powered to detect statistically significant differences. The other RCT, conducted in patients with amyotrophic lateral sclerosis, did not find significant improvements after high-frequency chest wall compression devices versus usual care in primary outcomes, in pulmonary function measures, or in most secondary outcomes. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
Clinical input obtained in 2008 supported the use of oscillatory devices to treat patients with cystic fibrosis and bronchiectasis, in certain situations. The most commonly mentioned clinical criteria were patients who failed or were intolerant of other methods of mucus clearance and patients who lacked caregivers to provide chest physical therapy. Thus, these devices may be considered medically necessary when chest physical therapy has failed, is unavailable, or is not tolerated by the patient.
The objective of this evidence review is to determine whether oscillatory devices improve the net health outcome in patients with cystic fibrosis and other respiratory disorders.
Use of an oscillatory positive expiratory pressure device may be considered medically necessary in individuals with hypersecretory lung disease (ie, produce excessive mucus) who have difficulty clearing the secretions and recurrent disease exacerbations.
High-frequency chest wall compression devices and intrapulmonary percussive ventilation devices may be considered medically necessary in individuals with cystic fibrosis or chronic diffuse bronchiectasis as determined by specific criteria (see Policy Guidelines section) (including chest computed tomography scan) when standard chest physical therapy has failed or standard chest physical therapy is unavailable or not tolerated. In considering the chest wall compression and intrapulmonary percussive ventilation devices, there should be demonstrated need for airway clearance. There should also be documented failure of standard treatments (ie, the patient has frequent severe exacerbations of respiratory distress involving inability to clear mucus despite standard treatment [chest physical therapy and, if appropriate, use of an oscillatory positive expiratory pressure device] or valid reasons why standard treatment cannot be performed, such as inability of the caregiver to perform it).
Other applications of high-frequency chest wall compression devices and intrapulmonary percussive ventilation devices, including, but not limited to, their use in individuals with cystic fibrosis or chronic diffuse bronchiectasis other than as specified above, their use as an adjunct to chest physical therapy, and their use in other lung diseases such as chronic obstructive pulmonary disease or respiratory conditions associated with neuromuscular disorders, are considered investigational.
For this policy, chronic diffuse bronchiectasis is defined by a daily productive cough for at least 6 continuous months or exacerbations more than 2 times per year requiring antibiotic therapy and confirmed by high-resolution or spiral chest computed tomography scan.
For the chest wall compression devices, a trial period to determine individual and family compliance may be considered. Those who appear to benefit most from the compression devices are adolescents and adults for whom, due to lifestyle factors, manual percussion and postural drainage may not be available.
A trial period may also be helpful because individuals' responses to different types of devices can vary; the types of devices should be considered as alternative, not equivalent, devices.
See the Codes table for details.
State or federal mandates (eg, Federal Employee Program) may dictate that certain U.S. Food and Drug Administration-approved devices, drugs, or biologics may not be considered investigational, and thus these devices may be assessed only by their medical necessity.
Oscillatory devices such as the Flutter device, the Vest® Airway Clearance System, and Percussionaire IPV® device have been primarily investigated as an alternative (not adjunct) to conventional chest physical therapy. Because published clinical data have not suggested that these devices are associated with an increased health benefit, their use would primarily represent a convenience to the patient. It is on this basis that they are considered not medically necessary (unless conventional chest physical therapy has failed or is unavailable).
Benefits are determined by the group contract, member benefit booklet, and/or individual subscriber certificate in effect at the time services were rendered. Benefit products or negotiated coverages may have all or some of the services discussed in this medical policy excluded from their coverage.
Oscillatory devices are designed to move mucus and clear airways; the oscillatory component can be intra- or extrathoracic. Some devices require the active participation of patients. They include oscillating positive expiratory pressure devices, such as Flutter and Acapella, in which the patient exhales multiple times through a device. The Flutter device is a small pipe-shaped, easily portable handheld device, with a mouthpiece at one end. It contains a high-density, stainless steel ball that rests in a plastic circular cone. During exhalation, the steel ball moves up and down, creating oscillations in expiratory pressure and airflow. When the oscillation frequency approximates the resonance frequency of the pulmonary system, the vibration of the airways occurs, resulting in loosening of mucus. The Acapella device is similar in concept but uses a counterweighted plug and magnet to create air flow oscillation.
Other airway clearance techniques also require active patient participation. For example, autogenic drainage and an active cycle breathing technique both involve a combination of breathing exercises performed by the patient. Positive expiratory pressure therapy requires patients to exhale through a resistor to produce positive expiratory pressures during a prolonged period of exhalation. It is hypothesized that the positive pressure supports the small airway such that the expiratory airflow can better mobilize secretions.
High-frequency chest wall oscillation devices (eg, the Vest Airway Clearance System) are passive oscillatory devices designed to provide airway clearance without active patient participation. The Vest Airway Clearance System provides high-frequency chest compression using an inflatable vest and an air-pulse generator. Large-bore tubing connects the vest to the air-pulse generator. The air-pulse generator creates pressure pulses that inflate and deflate the vest against the thorax, creating high-frequency chest wall oscillation and mobilization of pulmonary secretions.
All of these techniques may be alternatives to daily percussion and postural drainage in patients with cystic fibrosis, also known as chest physical therapy. Daily percussion and postural drainage need to be administered by a physical therapist or another trained adult in the home, often a parent if the patient is a child. The necessity for regular therapy can be particularly burdensome for adolescents or adults who lead independent lifestyles. Oscillatory devices can also potentially be used by patients with other respiratory disorders to promote bronchial secretion drainage and clearance, such as diffuse bronchiectasis and chronic obstructive pulmonary disease. Additionally, they could benefit patients with neuromuscular disease who have impaired cough clearance.
This evidence review addresses the outpatient use of oscillatory devices. This review does not address inpatient device use (eg, in the immediate postsurgical period).
Several oscillatory devices have been cleared for marketing by the U.S. Food and Drug Administration through the 510(k) process, including those listed in Table 1.
Device | Manufacturer | Clearance Date |
Flutter Mucus Clearance Device | Axcan Scandipharm (for marketing in the United States) | 1994 |
Vest Airway Clearance System | Hill-Rom | 1998 |
Acapella device | DHD Healthcare | 1999 |
RC Cornet® Mucus Clearing Device | PARI Respiratory Equipment | 1999 |
inCourage® System | RespirTech | 2005 |
Lung Flute® | Medical Acoustics LLC | 2006 |
Smartvest Airway Clearance System | Electromed | 2013 |
AerobiKA® oscillating PEP device | Trudell Medical | 2013 |
Vibralung® Acoustical Percussor | Westmed | 2014 |
The Vest Airway Clearance System | Hill-Rom | 2015 |
iPEP® system including PocketPEP® and vPEP® | D R Burton Healthcare | 2016 |
The Monarch™ Airway Clearance System | Hill-Rom | 2017 |
Pulsehaler™ | Respinova | 2021 |
PEP: positive expiratory pressure.U.S. Food and Drug Administration product codes: BYI, BYT.
This evidence review was created in November 1997 and has been updated regularly with searches of the PubMed database. The most recent literature update was performed through April 22, 2024 .
Evidence reviews assess the clinical evidence to determine whether the use of a technology improves the net health outcome. Broadly defined, health outcomes are length of life, quality of life (QOL), and ability to function-including benefits and harms. Every clinical condition has specific outcomes that are important to patients and to managing the course of that condition. Validated outcome measures are necessary to ascertain whether a condition improves or worsens; and whether the magnitude of that change is clinically significant. The net health outcome is a balance of benefits and harms.
To assess whether the evidence is sufficient to draw conclusions about the net health outcome of a technology, 2 domains are examined: the relevance and the quality and credibility. To be relevant, studies must represent 1 or more intended clinical use of the technology in the intended population and compare an effective and appropriate alternative at a comparable intensity. For some conditions, the alternative will be supportive care or surveillance. The quality and credibility of the evidence depend on study design and conduct, minimizing bias and confounding that can generate incorrect findings. The randomized controlled trial (RCT) is preferred to assess efficacy; however, in some circumstances, nonrandomized studies may be adequate. Randomized controlled trials are rarely large enough or long enough to capture less common adverse events and long-term effects. Other types of studies can be used for these purposes and to assess generalizability to broader clinical populations and settings of clinical practice.
Promotion of greater diversity and inclusion in clinical research of historically marginalized groups (e.g., People of Color [African-American, Asian, Black, Latino and Native American]; LGBTQIA (Lesbian, Gay, Bisexual, Transgender, Queer, Intersex, Asexual); Women; and People with Disabilities [Physical and Invisible]) allows policy populations to be more reflective of and findings more applicable to our diverse members. While we also strive to use inclusive language related to these groups in our policies, use of gender-specific nouns (e.g., women, men, sisters, etc.) will continue when reflective of language used in publications describing study populations.
The purpose of oscillatory positive expiratory pressure (PEP) therapy in individuals who have cystic fibrosis (CF) is to provide a treatment option that is an alternative to or an improvement on existing therapies.
The following PICO was used to select literature to inform this review.
The relevant population of interest is individuals with CF.
The therapy being considered is the application of oscillatory PEP. Oscillatory PEP devices are intended to be used primarily in the home setting by patients themselves.
The following therapy is currently being used: standard chest physical therapy.
The general outcomes of interest are reductions in respiratory symptoms due to airway restrictions caused by a mucous buildup in the lungs, QOL, hospitalizations, and medication use. Changes in outcomes over a minimum 3-month period should be considered meaningful.
Methodologically credible studies were selected using the following principles:
To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs;
In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies;
To assess long-term outcomes and adverse effects, single-arm studies that capture longer periods of follow-up and/or larger populations were sought;
A number of RCTs and a Cochrane systematic review of RCTs have evaluated oscillatory devices for treating patients with CF. The Cochrane review addressed a variety of oscillatory devices, was last updated by Morrison and Milroy (2020),1, and is summarized in Table 2. Outcomes included pulmonary function, sputum weight and volume, hospitalization rate, and QOL measures. Meta-analysis was limited due to the variety of devices, outcome measures, and lengths of follow-up used. Reviewers concluded that there was a lack of evidence supporting the superiority of oscillatory devices versus any other form of physical therapy, that one device was superior over another, and that there is a need for adequately powered RCTs with long-term follow-up.
Study | Dates | Trials | Participants | N (Range) | Design | Duration |
Morrison et al 20201, | Inception to July 2019 | 39 | Patients with cystic fibrosis | 1114 (4-166) | RCTs and controlled studies | 2 d to 2.8 y |
RCT: randomized controlled trial.
Representative recent RCTs follow. Trial characteristics and results are summarized in Tables 3 and 4. Gaps related to relevance, study design, and conduct are summarized in Tables 5 and 6.
Mcllwaine et al (2013) published an RCT comparing high-frequency chest wall oscillation (HFCWO) with PEP mask therapy.2,The primary outcome measure was the number of pulmonary exacerbations requiring an antibiotic. At the end of 1 year, patients in the PEP arm had a statistically significant lower incidence of pulmonary exacerbations requiring antibiotics compared with HFCWO group. The time to first pulmonary exacerbation was 220 days in the PEP group and 115 days in the HFCWO group (p=.02). There were no statistically significant differences in pulmonary measures, including the forced expiratory volume in 1 second (FEV1).
Sontag et al (2010) published a multicenter RCT that compared postural drainage, the Flutter device, and HFCWO.3, At study termination, patients had a final assessment; the length of participation ranged from 1.3 to 2.8 years. An intention-to-treat analysis found no significant differences between treatment groups in the modeled rate of decline for percent predicted FEV1 or forced vital capacity (FVC). The small sample size and high dropout rate limited the conclusions drawn from this trial.
Pryor et al (2010) evaluated 75 patients 16 years of age and older with CF from a single center in the U.K.4, Sixty-five (87%) of 75 patients completed the trial and were included in the analysis. Although the study was described as a noninferiority trial, it was not statistically analyzed as such. Instead, no statistically significant differences among the regimens in the primary outcome measure of FEV1 were construed as evidence for noninferiority.
The following study is not represented in the study tables within this review.
Radtke et al (2018) evaluated 15 adult patients with CF using the Flutter device with moderate-intensity interval cycling exercise to measure pulmonary diffusing capacity.5, The outcomes of interest included pulmonary function, sputum viscosity and volume, hospitalization rate, and QOL measures. The results yielded no differences in absolute changes in pulmonary diffusion capacity.
Study | Countries | Sites | Dates | Participants | Interventions | |
Active | Comparator | |||||
Mcllwaine et al (2013)2, | Canada | 12 | 2008 -2012 | Children with CF age >6 y (N=107) | HFCWO (n=56) | PEP mask therapy (n=51) |
Sontag et al (2010)3, | U.S. | 20 | 1999-2002 | Adults and children with CF (N=166) | 2 active Tx: flutter (n=58) and vest (n=57) | Postural drainage (n=58) |
Pryor et al (2010)4, | U.K. | 1 | NR | Patients with CF ≥16 y (N=75) | Cornet (n=15), Flutter (n=15), PEP (n=15), autogenic drainage (n=15) | Active cycle of breathing technique (n=15) |
CF: cystic fibrosis; HFCWO: high-frequency chest wall oscillation; NR: not reported; PEP: positive expiratory pressure; Tx: treatment.
Study | N | No. of PEs Requiring Antibiotics | Spirometry | Quality of Life |
Mcllwaine et al (2013)2, | 107 | Cannot confirm | Not applicable | |
HFCWO | Data not reported | Outcome not evaluated | ||
n | 96 | |||
Median | 2.00 | |||
Range | 1.00-3.00 | |||
Positive expiratory pressure | Data not reported | Outcome not evaluated | ||
n | 49 | |||
Median | 1.00 | |||
Range | 0.00-2.00 | |||
p | .007 | No difference | Not applicable | |
Sontag et al (2010)3, | ||||
Flutter | Outcome not evaluated | Data not reported | Outcome not evaluated | |
Vest | Outcome not evaluated | Data not reported | Outcome not evaluated | |
Postural drainage | Outcome not evaluated | Data not reported | Outcome not evaluated | |
p | No difference | |||
Pryor et al (2010)4, | 65 | Not applicable | Not applicable | |
Active cycle of breathing techniques | Outcome not evaluated | FEV1 at 0 mo: 2.01; FEV1 at 12 mo: 1.94 | Small improvement (0.7)a | |
Autogenic drainage | Outcome not evaluated | FEV1 at 0 mo: 2.68; FEV1 at 12 mo: 2.64 | Small improvement (0.5)a | |
Cornet | Outcome not evaluated | FEV1 at 0 mo: 1.93; FEV1 at 12 mo: 1.90 | No difference (<0.5)a | |
Flutter | Outcome not evaluated | FEV1 at 0 mo: 2.46; FEV1 at 12 mo: 2.43 | Moderate improvement (1.3)a | |
Positive expiratory pressure | Outcome not evaluated | FEV1 at 0 mo: 2.17; FEV1 at 12 mo: 2.02 | Small improvement (0.8)a | |
p | Not applicable | No difference | Not reported |
FEV1: forced expiratory volume in 1 second; HFCWO: high-frequency chest wall oscillation; PE: pulmonary exacerbations.a Minimal important differences in the Chronic Respiratory Questionnaire. A change of 0.5 represents a small difference in symptoms, 1.0 a moderate difference, and 1.5 a large difference
Study | Populationa | Interventionb | Comparatorc | Outcomesd | Duration of Follow-Upe |
Mcllwaine et al (2013)2, | |||||
Sontag et al (2010)3, | |||||
Pryor et al (2010)4, |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Population key: 1. Intended use population unclear; 2. Clinical context is unclear; 3. Study population is unclear; 4. Study population not representative of intended use.b Intervention key: 1. Not clearly defined; 2. Version used unclear; 3. Delivery not similar intensity as comparator; 4.Not the intervention of interest.c Comparator key: 1. Not clearly defined; 2. Not standard or optimal; 3. Delivery not similar intensity as intervention; 4. Not delivered effectively.d Outcomes key: 1. Key health outcomes not addressed; 2. Physiologic measures, not validated surrogates; 3. No CONSORT reporting of harms; 4. Not establish and validated measurements; 5. Clinical significant difference not prespecified; 6. Clinical significant difference not supported.e Follow-Up key: 1. Not sufficient duration for benefit; 2. Not sufficient duration for harms.
Study | Allocationa | Blindingb | Selective Reportingc | Data Completenessd | Powere | Statisticalf |
Mcllwaine et al (2013)2, | 3. Allocation concealment unclear | 1.Not blinded to treatment assignment | 1. Eighty-eight (82%) of 107 randomized patients completed the trial. Trial limitations were a nearly 20% dropout rate. | 4. Trial stopped early without enrolling expected number of patients and might have been underpowered to detect clinically significant differences between groups | ||
Sontag et al (2010)3, | 3. Allocation concealment unclear | 1.Not blinded to treatment assignment | 1. Dropout rates were high; trial ended early: 35 (60%), 16 (31%), and 5 (9%) patients withdrew from the postural drainage, Flutter, and Vest groups, respectively. Most common reasons for withdrawal after 60 days were moved or lost to follow-up (n=13) and lack of time (n=7). | 4. Trial ended earlier than planned | ||
Pryor et al (2010)4, | 3. Allocation concealment unclear | 1. Not blinded to treatment assignment | 1. Ten of 75 randomized patients were lost to follow-up |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Allocation key: 1. Participants not randomly allocated; 2. Allocation not concealed; 3. Allocation concealment unclear; 4. Inadequate control for selection bias.b Blinding key: 1. Not blinded to treatment assignment; 2. Not blinded outcome assessment; 3. Outcome assessed by treating physician.c Selective Reporting key: 1. Not registered; 2. Evidence of selective reporting; 3. Evidence of selective publication.dData Completeness key: 1. High loss to follow-up or missing data; 2. Inadequate handling of missing data; 3. High number of crossovers; 4. Inadequate handling of crossovers; 5. Inappropriate exclusions; 6. Not intent to treat analysis (per protocol for noninferiority trials).e Power key: 1. Power calculations not reported; 2. Power not calculated for primary outcome; 3. Power not based on clinically important difference f Statistical key: 1. Analysis is not appropriate for outcome type: (a) continuous; (b) binary; (c) time to event; 2. Analysis is not appropriate for multiple observations per patient; 3. Confidence intervals and/or p values not reported; 4.Comparative treatment effects not calculated.
A number of RCTs evaluating oscillatory devices have reported mixed findings and had limitations (eg, small sample sizes, large dropout rates). A systematic review identified 39 RCTs comparing oscillatory devices with other recognized airway clearance techniques; some were published only as abstracts. The study findings were not pooled due to heterogeneity in designs and outcome measures. The systematic review concluded that results from additional RCTs with adequate power and long-term follow-up would permit conclusions on the effect of oscillatory devices on outcomes for CF.
For individuals who have cystic fibrosis who receive oscillatory devices, the evidence includes randomized controlled trials (RCTs) and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. The RCTs reported mixed findings and limitations such as small sample sizes and large dropout rates. A systematic review identified 39 RCTs comparing oscillatory devices with other recognized airway clearance techniques; some were published only as abstracts. Reviewers could not pool findings due to heterogeneity in study designs and outcome measures and concluded that additional adequately powered RCTs with long-term follow-up would be needed to make conclusions about oscillatory devices for cystic fibrosis. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome. Medically necessary by Clinical input obtained in 2008.
Population Reference No. 1 Policy Statement | [X] MedicallyNecessary | [ ] Investigational |
The purpose of oscillatory PEP therapy in individuals who have bronchiectasis is to provide a treatment option that is an alternative to or an improvement on existing therapies.
The following PICO was used to select literature to inform this review.
The relevant population of interest is individuals with bronchiectasis.
The therapy being considered is the application of an oscillatory PEP. Oscillatory PEP devices are intended to be used primarily in the home setting by patients themselves.
The following therapy is currently being used: standard chest physical therapy.
The general outcomes of interest are reductions in respiratory symptoms due to airway restrictions (eg, pulmonary exacerbations), QOL, hospitalizations, and medication use. Changes in outcomes over a minimum 3-month period should be considered meaningful.
Methodologically credible studies were selected using the following principles:
To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs;
In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies;
To assess long-term outcomes and adverse effects, single-arm studies that capture longer periods of follow-up and/or larger populations were sought;
Studies with duplicative or overlapping populations were excluded.
Lee et al (2015) published a Cochrane review of airway clearance techniques for treating bronchiectasis, which is summarized in Table 7.6, Of 7 RCTs included, 6 were crossover trials. Five trials used a PEP device, 1 used HFCWO, and 1 used postural drainage. Reviewers did not pool study findings due to heterogeneity among studies. Primary outcomes of interest were pulmonary exacerbations, hospitalizations for bronchiectasis, and QOL.
Study | Dates | Trials | Participants | N (Range) | Design | Duration |
Lee et al (2015)6, | 1966-2015 | 7 RCTs | Adults and children diagnosed with bronchiectasis based on plain-film chest radiography, bronchography, high-resolution computed tomography, or physician diagnosis | 1107 (8-37) | 1 RCT, 6 crossover RCTs | Immediate (within 24 h) and "long-term" (>24 h) |
RCT: randomized controlled trial.
Representative recent RCTs follow. Trial characteristics and results are summarized in Tables 8 and 9. Gaps related to relevance, study design, and conduct are summarized in Tables 10 and 11.
Murray et al (2009) reported on a crossover study with 20 patients. The number of exacerbations did not differ statistically at 12 weeks.7, Cough-related QOL was significantly better after 12 weeks of any airway clearance technique compared with no airway clearance. Cochrane reviewers noted that the study was not blinded and that patient-reported QOL measures may have been subject to bias.
Herrero-Cortina et al (2016) reported on a crossover RCT with 31 patients.8, The interventions were temporary PEP, autogenic drainage, and slow expiration with the glottis opened in the lateral position. There were no significant differences among treatments in the mean sputum clearance during the 24-hour period after each intervention, cough severity (measured using the total Leicester Cough Questionnaire [LCQ] score), or lung function measures (eg, FEV1).
Livnat et al (2021) conducted a randomized trial in 51 patients with bronchiectasis that compared autogenic drainage and oscillating PEP for daily airway clearance.9, Patients who had not previously performed airway clearance were included. After 4 weeks, the primary outcome (lung clearance index, calculated as the cumulative expired volume during the washout phase divided by the functional residual capacity) and FEV1 did not differ between groups. Change in sputum quantity from randomization to study end did not differ between groups. The rate of exacerbations was not described, but some QOL measures improved throughout the study in both groups.
Study | Countries | Sites | Dates | Participants | Interventions | |
Active | Comparator | |||||
Murray et al (2009)7, | U.K. | 1 | NR | Patients radiologically diagnosed with bronchiectasis (N=20) | Acapella Choice (n=20) | No chest physical therapy (n=20) |
Herrero-Cortina et al (2016)8, | Spain | 1 | 2010-2013 | Patients radiologically diagnosed with bronchiectasis (N=31) | Slow expiration with glottis opened in lateral posture (n=31) and temporary PEP (n=31) | Autogenic drainage (n=31) |
Livnat et al (2021)9, | Israel | 1 | 2017-2019 | Patients radiologically diagnosed with bronchiectasis (N=51) | Aerobika (n=24) | Autogenic drainage (n=25) |
NR: not reported; PEP: positive expiratory pressure.
Study | Total LCQ Score Difference | 24-h Sputum Volume Difference, mL | No. of Exacerbations |
Median (IQR) | Median (IQR) | ||
Murray et al (2009)7, | N=20 | N=20 | Not applicable |
Acapella | 1.3 (-0.17 to 3.25) | 2 (0 to 6) | 5 |
No Acapella | 0 (-1.5 to 0.5) | -1 (-5 to 0) | 7 |
p | .002 | .02 | .48 |
Herrero-Cortina et al (2016)8, | |||
Autogenic drainage | 0.5 (0.1 to 0.5);.01 | -1.4 (5.1 to 1.2) | Not studied |
ELTGOL | 0.9 (0.5 to 2.1);.001 | -1.6 (-4.8 to 1.0) | Not studied |
TPEP | 0.4 (0.1 to 1.2);.04 | -2.5 (-8.6 to 0.1) | Not studied |
p | See above | .01 | Not applicable |
Livnat et al (2021)9, | |||
Aerobika | Not studied | -10 | Not studied |
Autogenic drainage | Not studied | -2.2 | Not studied |
p | Not applicable | .386 | Not applicable |
ELTGOL: expiration with glottis opened in lateral posture; IQR: interquartile range; LCQ: Leicester Cough Questionnaire; TPEP: temporary positive expiratory pressure.
Study | Populationa | Interventionb | Comparatorc | Outcomesd | Duration of Follow-Upe |
Murray et al (2009)7, | |||||
Herrero-Cortina et al (2016)8, | 1, 2. 24-h follow-up is not enough | ||||
Livnat et al (2021)9, | 1. No data on exacerbations |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Population key: 1. Intended use population unclear; 2. Clinical context is unclear; 3. Study population is unclear; 4. Study population not representative of intended use.b Intervention key: 1. Not clearly defined; 2. Version used unclear; 3. Delivery not similar intensity as comparator; 4.Not the intervention of interest.c Comparator key: 1. Not clearly defined; 2. Not standard or optimal; 3. Delivery not similar intensity as intervention; 4. Not delivered effectively.d Outcomes key: 1. Key health outcomes not addressed; 2. Physiologic measures, not validated surrogates; 3. No CONSORT reporting of harms; 4. Not establish and validated measurements; 5. Clinical significant difference not prespecified; 6. Clinical significant difference not supported.e Follow-Up key: 1. Not sufficient duration for benefit; 2. Not sufficient duration for harms.
Study | Allocationa | Blindingb | Selective Reporting c | Data Completeness d | Power e | Statisticalf |
Murray et al (2009)7, | 3. Allocation concealment unclear | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment 3. Outcome assessed by treating physician | 3. Power not based on clinically important difference | |||
Herrero-Cortina et al (2016)8, | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment 3. Outcome assessed by treating physician | 1. Power calculations not reported 2. Power not calculated for primary outcome 3. Power not based on clinically important difference | ||||
Livnat et al (2021)9, | 1. Not blinded to treatment assignment (participants) |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment,a Allocation key: 1. Participants not randomly allocated; 2. Allocation not concealed; 3. Allocation concealment unclear; 4. Inadequate control for selection bias.b Blinding key: 1. Not blinded to treatment assignment; 2. Not blinded outcome assessment; 3. Outcome assessed by treating physician.c Selective Reporting key: 1. Not registered; 2. Evidence of selective reporting; 3. Evidence of selective publication.dData Completeness key: 1. High loss to follow-up or missing data; 2. Inadequate handling of missing data; 3. High number of crossovers; 4. Inadequate handling of crossovers; 5. Inappropriate exclusions; 6. Not intent to treat analysis (per protocol for noninferiority trials).e Power key: 1. Power calculations not reported; 2. Power not calculated for primary outcome; 3. Power not based on clinically important difference f Statistical key: 1. Analysis is not appropriate for outcome type: (a) continuous; (b) binary; (c) time to event; 2. Analysis is not appropriate for multiple observations per patient; 3. Confidence intervals and/or p values not reported; 4.Comparative treatment effects not calculated.
A 2015 systematic review identified 7 small RCTs assessing several types of oscillatory devices; only 1 reported the clinically important outcomes of exacerbations or hospitalizations. Three reported on QOL, and trial findings were mixed. A 2016 crossover RCT did not find a significant benefit of temporary PEP compared with other airway clearance techniques.
For individuals who have bronchiectasis who receive oscillatory devices, the evidence includes RCTs and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. A 2015 systematic review identified 7 small RCTs on several types of oscillatory devices; only 1 reported the clinically important outcomes of exacerbations or hospitalizations. Only 3 RCTs reported on quality of life, and findings were mixed. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome. Medically necessary by Clinical input obtained in 2008.
Population Reference No. 2 Policy Statement | [X] MedicallyNecessary | [ ] Investigational |
The purpose of oscillatory PEP therapy in individuals who have chronic obstructive pulmonary disease (COPD) is to provide a treatment option that is an alternative to or an improvement on existing therapies.
The following PICO was used to select literature to inform this review.
The relevant population of interest is individuals with COPD.
The therapy being considered is the application of an oscillatory PEP. Oscillatory PEP devices are intended to be used primarily in the home setting by patients themselves.
The following therapy is currently being used: standard therapy.
The general outcomes of interest are reductions in respiratory symptoms due to airway restrictions (eg, pulmonary exacerbations), QOL, hospitalizations, and medication use. Changes in outcomes over a minimum 3-month period should be considered meaningful.
Methodologically credible studies were selected using the following principles:
To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs;
In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies;
To assess long-term outcomes and adverse effects, single-arm studies that capture longer periods of follow-up and/or larger populations were sought;
Studies with duplicative or overlapping populations were excluded.
Systematic reviews have evaluated studies of airway clearance techniques in patients with COPD.10,11,12, Two early reviews addressed various techniques (ie, they were not limited to studies on oscillatory devices) while the most recent review was specific to oscillatory devices. These are summarized in Table 12. Studies included in the systematic reviews were mostly small and reviewers noted that the quality of evidence was generally poor. The meta-analysis conducted by Alghamdi et al found oscillatory PEP reduced exacerbations (odds ratio, 0.37; 95% confidence interval [CI], 0.19 to 0.72) and improved 6-minute walk distance (mean difference, 49.8 m; 95% CI, 14.2 to 85.5 m), but the authors also noted the need for higher-quality studies.13,
Study | Dates | Trials | Participants | N (Range) | Design | Duration |
Ides et al (2011)10, | 1980-2008 | 26 | Patients with COPD | 659 (7-58) | Not reported | Unclear |
Osadnik et al (2012)11, | Inception to 2009 (PEDro) or 2011 (CAGR) | 28 | Participants with investigator-defined COPD, emphysema or chronic bronchitis | 907 (5-96) | RCTs (parallel and crossover) | 24 h to >8 wk |
Alghamdi et al (2020)13, | Inception to March 2020 | 8 | Patients with COPD | 381 (15-120) | RCTs and crossover | 5 d to 2 y |
CAGR: Cochrane Airways Group Specialised Register of trials; COPD: chronic obstructive pulmonary disease; PEDro: Physiotherapy Evidence Database; RCT: randomized controlled trial.
Representative recent RCTs follow. Trial characteristics and results are summarized in Tables 13 and 14. Gaps related to relevance, study design and conduct are summarized in Tables 15 and 16.
Chakrovorty et al (2011) reported results of a crossover RCT among patients with moderate-to-severe COPD and mucus hypersecretion.14, Patients received HFCWO or conventional treatment in random order, for 4 weeks, with a 2-week washout period between treatments. The primary outcome was QOL as measured using the St. George's Respiratory Questionnaire (SGRQ). Only 1 of 4 dimensions of the SGRQ (the symptom dimension) improved after HFCWO compared with baseline, with a decrease in mean score from 72 to 64 (p=.02). None of the 4 SGRQ dimensions improved after conventional treatment. There were no significant pre- to posttreatment differences in secondary outcomes (eg, FEV1, FVC).
Svenningsen et al (2016) reported on the results of an unblinded, industry-funded, randomized crossover study.15, Each intervention period lasted 21 to 28 days. In the nonsputum producers, scores differed significantly only on the Patient Evaluation Questionnaire total score. In patients who were sputum-producers at baseline, pre- versus post-PEP scores differed significantly for FVC, 6-minute walk distance, SGRQ total score, and the Patient Evaluation Questionnaire ease of bringing up sputum and patient global assessment subscales. It is unclear if the interventions were clinically meaningful. The crossover studies had similar limitations including no between-group comparisons (ie, outcomes after oscillatory device use vs. the control intervention), lack of intention-to-treat analysis, and short-term follow-up (immediate posttreatment period).
Goktalay et al (2013) reported on the results of a parallel-group RCT.16, Patients were randomized to 5 days of treatment with medical therapy plus HFCWO (n=25) or medical therapy only (n=25). At day 5, outcomes including FEV1, modified Medical Research Council dyspnea scale scores, and the 6-minute walk distance, did not differ significantly between groups. This short-term trial included hospitalized patients who might differ from COPD patients treated on an outpatient basis.
Alghamdi et al (2023) compared the Acapella device to usual care in patients with stable COPD (N=122).13, The primary outcome was the change from baseline in LCQ score. Results demonstrated significant improvement in LCQ scores with the use of Acapella compared to usual care.
Study | Countries | Sites | Dates | Participants | Interventions | |
Active | Comparator | |||||
Chakrovorty et al (2011)14, | U.K. | 1 | NR | Patients with at least 1 COPD exacerbation with FEV1 <0.8, FEV1/FVC <0.7, and a daily wet sputum volume of >25 mL (N=38) (female, n=8; male, n=30) | SmartVest Airway Clearance System (n=22) | No SmartVest Airway Clearance System (n=22) |
Svenningsen et al (2016)15, | Canada | 1 | NR | COPD patients self-identified as sputum-producers or non-sputum-producers (N=32) (female, n=13; male, n=14) | Oscillatory PEP (AerobiKA device) (n=27) | No oscillatory PEP (n=27) |
Goktalay et al (2013)16, | Turkey | 1 | 2009-2011 | Patients with stage 3 or 4 COPD hospitalized for COPD exacerbations (N=50) (female, n=1; male, n=49) | HFCWO plus medical Tx (n=25) | Medical Tx only (n=25) |
Alghamdi et al (2023)13, | NR | 1 | 2020-2021 | Stable COPD patients self-identified as sputum producers every day or most days (N=122) (female, n=49; male n=73) | Oscillatory PEP (Acapella) (n=61) | Usual care, including active cycle of breathing technique (n=61) |
COPD: chronic obstructive pulmonary disease; FEV1: forced expiratory volume in 1 second; FVC: forced vital capacity; HFCWO: high-frequency chest wall oscillation; NR: not reported; PEP: positive expiratory pressure; Tx: treatment.
Study | SGRO Total Scores | BODE Index | LCQ score change from baseline |
Chakrovorty et al (2011)14, | |||
SmartVest | Baseline: 63; End of treatment: 60 | Not assessed | |
No SmartVest | Baseline: 62; End of treatment:62 | Not assessed | |
p | NS | Not applicable | |
Svenningsen et al (2016)15, | |||
Oscillatory positive expiratory pressure | Sputum-producers: 40 (12); Non-sputum-producers: 36 | Not assessed | |
Control | Sputum-producers: 49; Non-sputum-producers: 35 | Not assessed | |
p | .01 (sputum-producers);.64 (non-sputum-producers) | Not applicable | |
Goktalay et al (2013)16, | |||
HFCWO plus medical treatment | Not assessed | Day 0: 7.72; Day 3: 7.00; Day 5: 6.44 | |
Medical treatment only | Not assessed | Day 0: 7.72; Day 3: 7.48; Day 5: 7.24 | |
p | Not applicable | Uninterpretable | |
Alghamdi et al (2023)13, | |||
Oscillatory positive expiratory pressure | 1.54 (0.33 to 2.18) | ||
Usual care | 0.51 (0.34 to 1.89) | ||
MD (95% CI); p | 1.03 (0.71 to 2.10);.03 |
BODE: body mass index, airflow obstruction, dyspnea, and exercise; CI: confidence interval; HFCWO: high-frequency chest wall oscillation; LCQ: Leicester Cough Questionnaire; MD: mean difference; NS: not significant; SGRO: St George's Respiratory Questionnaire.
Study | Populationa | Interventionb | Comparatorc | Outcomesd | Duration of Follow-Upe |
Chakrovorty et al (2011)14, | |||||
Svenningsen et al (2016)15, | |||||
Goktalay et al (2013)16, | 1. Not sufficient duration for benefits (short-term follow-up for 5 d) | ||||
Alghamdi et al (2023)13, |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Population key: 1. Intended use population unclear; 2. Clinical context is unclear; 3. Study population is unclear; 4. Study population not representative of intended use.b Intervention key: 1. Not clearly defined; 2. Version used unclear; 3. Delivery not similar intensity as comparator; 4.Not the intervention of interest.c Comparator key: 1. Not clearly defined; 2. Not standard or optimal; 3. Delivery not similar intensity as intervention; 4. Not delivered effectively.d Outcomes key: 1. Key health outcomes not addressed; 2. Physiologic measures, not validated surrogates; 3. No CONSORT reporting of harms; 4. Not establish and validated measurements; 5. Clinical significant difference not prespecified; 6. Clinical significant difference not supported.e Follow-Up key: 1. Not sufficient duration for benefit; 2. Not sufficient duration for harms.
Study | Allocationa | Blindingb | Selective Reporting c | Data Completeness d | Power e | Statisticalf |
Chakrovorty et al (2011)14, | 3. Allocation concealment unclear | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment 3. Outcome assessed by treating physician | 1. High loss to follow-up or missing data: 8 out of 30 withdrew due to COPD exacerbations | 2. Power not calculated for primary outcome | ||
Svenningsen et al (2016)15, | 3. Allocation concealment unclear | 1. Not blinded to treatment assignment | 1. High loss to follow-up or missing data: 16% withdrew from trial | 2. Power not calculated for primary outcome | ||
Goktalay et al (2013)16, | 1. Participants not randomly allocated 2. Allocation not concealed | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment 3. Outcome assessed by treating physician | 1. Power calculations not reported 2. Power not calculated for primary outcome 3. Power not based on clinically important difference | |||
Alghamdi et al (2023)13, | 1. Not blinded to treatment assignment | 1. High loss to follow-up or missing data: 15% lost to follow-up and 9% with no follow-up data for objective monitoring |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.COPD: chronic obstructive pulmonary disease.a Allocation key: 1. Participants not randomly allocated; 2. Allocation not concealed; 3. Allocation concealment unclear; 4. Inadequate control for selection bias.b Blinding key: 1. Not blinded to treatment assignment; 2. Not blinded outcome assessment; 3. Outcome assessed by treating physician.c Selective Reporting key: 1. Not registered; 2. Evidence of selective reporting; 3. Evidence of selective publication.dData Completeness key: 1. High loss to follow-up or missing data; 2. Inadequate handling of missing data; 3. High number of crossovers; 4. Inadequate handling of crossovers; 5. Inappropriate exclusions; 6. Not intent to treat analysis (per protocol for noninferiority trials).e Power key: 1. Power calculations not reported; 2. Power not calculated for primary outcome; 3. Power not based on clinically important difference f Statistical key: 1. Analysis is not appropriate for outcome type: (a) continuous; (b) binary; (c) time to event; 2. Analysis is not appropriate for multiple observations per patient; 3. Confidence intervals and/or p values not reported; 4.Comparative treatment effects not calculated.
Only a few controlled studies have evaluated oscillatory devices for the treatment of COPD, and they tended to use intention-to-treat analysis and between-group comparisons. The published studies reported mixed findings and did not support the use of oscillatory devices in patients with COPD.
For individuals who have chronic obstructive pulmonary disease (COPD) who receive oscillatory devices, the evidence includes RCTs and systematic reviews. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. Only a few controlled studies have evaluated oscillatory devices for the treatment of COPD, and they tend to have small sample sizes, short follow-up periods, and limitations in their analyses (eg, lack of intention-to-treat analysis and between-group comparisons). Moreover, the published studies reported mixed findings and did not consistently support the use of oscillatory devices in this population. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
Population Reference No. 3 Policy Statement | [ ] MedicallyNecessary | [X] Investigational |
The purpose of oscillatory PEP therapy in individuals who have respiratory conditions related to neuromuscular disorders is to provide a treatment option that is an alternative to or an improvement on existing therapies.
The following PICO was used to select literature to inform this review.
The relevant population of interest is individuals with respiratory conditions related to neuromuscular disorders.
The therapy being considered is the application of an oscillatory PEP. Oscillatory PEP devices are intended to be used primarily in the home setting by patients themselves.
The following therapy is currently being used: standard therapy.
The general outcomes of interest are reductions in respiratory symptoms due to airway restrictions (eg, pulmonary exacerbations), QOL, hospitalizations, and medication use. Changes in outcomes over a minimum 3-month period should be considered meaningful.
Methodologically credible studies were selected using the following principles:
To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs;
In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies;
To assess long-term outcomes and adverse effects, single-arm studies that capture longer periods of follow-up and/or larger populations were sought;
Studies with duplicative or overlapping populations were excluded.
A Cochrane review by Winfield et al (2014) evaluated the nonpharmacologic management of respiratory morbidity in children with severe global developmental delay treated with airway clearance techniques.17, Reviewers included RCTs and nonrandomized comparative studies. They identified 3 studies on HFCWO (1 RCT, 2 pre-post) and one on PEP (pre-post), with sample sizes from 15 and 28 patients. As a result of heterogeneity, a meta-analysis was not conducted. The reviewis summarized in Table 17.
Study | Dates | Trials | Participants | N (Range) | Design | Duration |
Winfield et al (2014)17, | Inception to Nov 2013 | 15 | Children up to 18 y with a diagnosis of severe neurologic impairment and respiratory morbidity | Not reported | RCTs and nonrandomized comparative studies | Unclear |
RCT: randomized controlled trial.
Representative recent RCTs follow. Trial characteristics and results are summarized in Tables 18 and 19. Gaps related to relevance, study design and conduct are summarized in Tables 20 and 21.
Yuan et al (2010) reported results of a parallel-arm RCT.18, Both groups were instructed to perform the assigned treatment for 12 minutes, 3 times a day for the study period (mean, 5 months). There were no statistically significant differences between groups on primary outcomes. No therapy-related adverse events were reported in either group.
Lange et al (2006) reported on the results of a parallel-arm RCT in adults with amyotrophic lateral sclerosis.19, Patients were randomized to 12 weeks of HCFWO or usual care. There were no statistically significant between-group differences in pulmonary measures (FVC predicted, capnography, oxygen saturation, or peak expiratory flow). There was also no significant difference in the amyotrophic lateral sclerosis Functional Rating Scale respiratory subscale score (worsening) at 12 weeks. Of symptoms assessed as secondary outcomes, there was significantly less breathlessness and night cough in the HCFWO group than in the usual care group, and groups did not differ significantly on other symptoms, including the noise of breathing, suction frequency, suction amount, day cough, and nocturnal symptoms.
Study | Countries | Sites | Dates | Participants | Interventions | |
Active | Comparator | |||||
Yuan et al (2010)18, | U.S. | 1 | NR | Patients with cerebral palsy or neuromuscular disease attending a pediatric pulmonary clinic (N=28) (Hispanic, n=9; White, n=7; Asian, n=4; African American, n=2; Pacific Islander, n=1) | HCFWO (n=12) | Standard chest physical therapy (n=11) |
Lange et al (2006)19, | U.S. | 6 | NR | Adults with amyotrophic lateral sclerosis (N=46) | HCFWO (n=22) | No treatment (n=24) |
HFCWO: high-frequency chest wall oscillation; NR: not reported.
Study | Hospitalization/IV Antibiotics | TDI (proportion showing worsening) |
Yuan et al (2010)18, | N=23 | |
HCFWO | 0/12 | Not assessed |
Standard chest physical therapy | 4/11 | Not assessed |
p | .09 | Not applicable |
Lange et al (2006)19, | - | N=18 |
HCFWO | Not assessed | Functional impairment: 27.8%; Magnitude of task: 38.9%; Magnitude of effort: 27.8% |
No treatment | Not assessed | Functional impairment: 43.8%; Magnitude of task: 50%; Magnitude of effort: 56.2% |
p | Not applicable | Functional impairment:.331; Magnitude of task:.515; Magnitude of effort:.092 |
HFCWO: high- frequency chest wall oscillation; IV: intravenous; TDI: Transitional Dyspnea Index.
Study | Populationa | Interventionb | Comparatorc | Outcomesd | Duration of Follow-Upe |
Yuan et al (2010)18, | |||||
Lange et al (2006)19, |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Population key: 1. Intended use population unclear; 2. Clinical context is unclear; 3. Study population is unclear; 4. Study population not representative of intended use.b Intervention key: 1. Not clearly defined; 2. Version used unclear; 3. Delivery not similar intensity as comparator; 4.Not the intervention of interest.c Comparator key: 1. Not clearly defined; 2. Not standard or optimal; 3. Delivery not similar intensity as intervention; 4. Not delivered effectively.d Outcomes key: 1. Key health outcomes not addressed; 2. Physiologic measures, not validated surrogates; 3. No CONSORT reporting of harms; 4. Not establish and validated measurements; 5. Clinical significant difference not prespecified; 6. Clinical significant difference not supported.e Follow-Up key: 1. Not sufficient duration for benefit; 2. Not sufficient duration for harms.
Study | Allocationa | Blindingb | Selective Reporting c | Data Completeness d | Power e | Statisticalf |
Yuan et al 2010)18, | 1. Allocation concealment unclear | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment (except chest X-rays) 3. Outcome assessed by treating physician | 1. High loss to follow-up or missing data 12% missing data and all in treatment group | 1, 2, 3. Trial was exploratory and was not powered to detect statistically significant findings of the primary outcomes | ||
Lange et al (2006)19, | 1. Allocation not concealed | 1. Not blinded to treatment assignment 2. Not blinded outcome assessment 3. Outcome assessed by treating physician | 1. High loss to follow-up or missing data 15% missing data at 12 wk | 2. Power not calculated for primary outcome 3. Power not based on clinically important difference |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment.a Allocation key: 1. Participants not randomly allocated; 2. Allocation not concealed; 3. Allocation concealment unclear; 4. Inadequate control for selection bias.b Blinding key: 1. Not blinded to treatment assignment; 2. Not blinded outcome assessment; 3. Outcome assessed by treating physician.c Selective Reporting key: 1. Not registered; 2. Evidence of selective reporting; 3. Evidence of selective publication.dData Completeness key: 1. High loss to follow-up or missing data; 2. Inadequate handling of missing data; 3. High number of crossovers; 4. Inadequate handling of crossovers; 5. Inappropriate exclusions; 6. Not intent to treat analysis (per protocol for noninferiority trials).e Power key: 1. Power calculations not reported; 2. Power not calculated for primary outcome; 3. Power not based on clinically important difference f Statistical key: 1. Analysis is not appropriate for outcome type: (a) continuous; (b) binary; (c) time to event; 2. Analysis is not appropriate for multiple observations per patient; 3. Confidence intervals and/or p values not reported; 4.Comparative treatment effects not calculated.
Two RCTs and a systematic review have evaluated oscillatory devices for the treatment of respiratory conditions in neuromuscular disorders. One RCT was not powered to detect statistical significance. The other, conducted in amyotrophic lateral sclerosis patients, did not find statistically significant improvement after HCFWO compared with usual care for the primary outcomes (pulmonary function measures) or most secondary outcomes.
For individuals who have respiratory conditions related to neuromuscular disorders who receive oscillatory devices, the evidence includes 2 RCTs and a systematic review. Relevant outcomes are symptoms, quality of life, hospitalizations, and medication use. One of the RCTs was not powered to detect statistically significant differences. The other RCT, conducted in patients with amyotrophic lateral sclerosis, did not find significant improvements after high-frequency chest wall compression devices versus usual care in primary outcomes, in pulmonary function measures, or in most secondary outcomes. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
Population Reference No. 4 Policy Statement | [ ] MedicallyNecessary | [X] Investigational |
The purpose of the following information is to provide reference material. Inclusion does not imply endorsement or alignment with the evidence review conclusions.
While the various physician specialty societies and academic medical centers may collaborate with and make recommendations during this process, through the provision of appropriate reviewers, input received does not represent an endorsement or position statement by the physician specialty societies or academic medical centers, unless otherwise noted.
In response to requests, input was received from 2 academic medical centers while this policy was under review in 2008. Input indicated the available studies demonstrated that these oscillatory devices are comparable with chest physical therapy for cystic fibrosis and bronchiectasis. The most commonly mentioned clinical criteria were patients who failed or were intolerant of other methods of mucus clearance and patients who lacked caregivers to provide chest physical therapy. Input did not support the use of oscillatory devices for treatment of chronic obstructive pulmonary disease.
Guidelines or position statements will be considered for inclusion in ‘Supplemental Information’ if they were issued by, or jointly by, a US professional society, an international society with US representation, or National Institute for Health and Care Excellence (NICE). Priority will be given to guidelines that are informed by a systematic review, include strength of evidence ratings, and include a description of management of conflict of interest.
In 2006, the guidelines from the American College of Chest Physicians recommended (level of evidence: low) that, in patients with cystic fibrosis, devices designed to oscillate gas in the airway, either directly or by compressing the chest wall, can be considered as an alternative to chest physical therapy.20,
A 2018 document from the American College of Chest Physicians recommends that airway clearance strategies in children and adults with productive cough due to bronchiectasis related to any cause be individualized to the patient (ungraded, consensus statement).21,
In 2009, the Cystic Fibrosis Foundation published guidelines on airway clearance therapies based on a systematic review of evidence.22, The Foundation recommended airway clearance therapies for all patients with cystic fibrosis but stated that no therapy had been demonstrated to be superior to others (level of evidence: fair; net benefit: moderate; grade of recommendation: B).
Not applicable.
There is no national coverage determination. In the absence of a national coverage determination, coverage decisions are left to the discretion of local Medicare carriers.
Some currently ongoing trials that might influence this review are listed in Table 22.
NCT No. | Trial Name | Planned Enrollment | Completion Date |
Ongoing | |||
NCT05548036 | A Feasibility Randomised Control Trial (RCT) of Aerobika TM Verses Active Cycle of Breathing Technique (ACBT) in People With Chronic Obstructive Pulmonary Disease (COPD) (TIPTOP) | 120 | Apr 2024 |
Completed | |||
NCT05034900 | Does Addition of Oscillatory Positive Expiratory Pressure (OPEP) Device to a Chest Physiotherapy Program Provide Further Health Benefits in Children With Bronchiectasis? | 42 | Sept 2022 |
NCT04271969 | Clinical Effectiveness Of High Frequency Chest Wall Oscillation (HFCWO) In A Bronchiectasis Population | 125 | Dec 2023 |
NCT: national clinical trial.
Codes | Number | Description |
CPT | No Code | |
HCPCS | A7025 | High frequency chest wall oscillation system vest, replacement for use with patient-owned equipment, each |
| A7026 | High frequency chest wall oscillation system hose, replacement for use with patient-owned equipment, each |
| E0480 | Percussor, electric or pneumatic, home model |
| E0481 | Intrapulmonary percussive ventilation system and related accessories |
| E0483 | High frequency chest wall oscillation air-pulse generator system (includes hoses and vest), each |
| E0484 | Oscillatory positive expiratory pressure device, non-electric, any type, each |
| S8185 | Flutter device |
ICD-10-CM | E84.0-E84.9 | Cystic fibrosis code range |
| J47.1-J47.9 | Bronchiectasis code range |
ICD-10-PCS | | ICD-10-PCS codes are only used for inpatient services. There are no ICD-10-PCS codes for devicesand there is no is no specific ICD-10-PCS code for this therapy. |
| F07C6ZZ | Physical rehabilitation and diagnostic audiology, rehabilitation, motor treatment, respiratory system- whole body, therapeutic exercise |
Type of service | Pulmonary | |
Place of service | Home | |
As per correct coding guidelines.
Date | Action | Description |
---|---|---|
09/09/2024 | Annual Review | Policy updated with literature review through April 22, 2024; no references added. Policy statements unchanged. The need for an active policy was affirmed. Summary of Key Trials Table (22) was updated. |
07/12/2024 | Annual Review | No changes. |
07/05/2023 | Annual Review | Policy updated with literature review through April 19, 2023; reference added. Policy statements unchanged. |
07/06/2022 | Annual Review | Policy updated with literature review through April 18, 2022; references added. Minor editorial refinements made to policy statements; intent unchanged. |
07/19/2021 | Annual Review | Policy updated with literature review through April 19, 2021; references added. Policy statements unchanged |
11/09/2020 | Annual Review | Policy statement unchanged. |
07/16/2020 | Annual Review | Policy statement unchanged. |
07/17/2019 | Policy reviewed | Medical necessity statement changed due to clinical input |
06/09/2017 | Policy reviewed | Policy updated with literature review through April 25, 2017; reference 9 added. ‘Not medically necessary' statement removed and "patients with cystic fibrosis or chronic diffuse bronchiectasis other than as specified above" added to the investigational statement. |
06/15/2016 | Policy reviewed | Policy unchanged |
02/12/2015 | Policy reviewed | Policy updated with literature review through December 15, 2014. Reference 1 added. Policy statements unchanged. |
02/13/2014 | Policy reviewed | Policy updated with literature review through December 20, 2013. In first 2 medically necessary statements, Flutter or Flutter and Acapella changed to oscillatory positive expiratory pressure device. References 2, 7-9, and 13 added; other references renumbered or removed. In second policy statement, "standard chest physiotherapy treatment" changed to "standard treatment". |
12/11/2013 | Policy reviewed | Policy unchanged |
05/02/2012 | Policy reviewed | Policy unchanged |
02/09/2012 | Policy changed | ICD-10 added |
07/10/2009 | Policy created | New policy |