Medical Policy
Policy Num: 02.001.048
Policy Name: Bioimpedance Devices for Detection and Management of Lymphedema
Policy ID: [02.001.048] [Ac / B / M- / P-] [2.01.82]
Last Review: February 10, 2025
Next Review: February 20, 2026
Related Policies:
08.001.029 - Pneumatic Compression Pumps for Treatment of Lymphedema and Venous Ulcers
07.001.119 - Surgical Treatments for Breast Cancer-Related Lymphedema
07.001.164 Liposuction for Lipedema and Lymphedema
Population Reference No. | Populations | Interventions | Comparators | Outcomes |
1 | Individuals:
| Interventions of interest are:
| Comparators of interest are:
| Relevant outcomes include:
|
Secondary lymphedema may develop following treatment for breast cancer. Bioimpedance, which uses resistance to electrical current to compare the composition of fluid compartments, could be used as a tool to diagnose lymphedema.
For individuals who have known or suspected lymphedema who receive bioimpedance spectroscopy, the evidence includes systematic reviews, 1 randomized controlled trial (RCT), 1 prospective comparative observational study, and multiple uncontrolled observational studies. Relevant outcomes are test validity, symptoms, and quality of life. Diagnostic accuracy studies have found a poor correlation between bioimpedance analysis and the reference standard (volume displacement or circumferential measurement). Results from the PREVENT RCT comparing bioimpedance with standard tape measure following treatment for breast cancer have been published. At a median follow-up of 32.9 months, BIS patients triggered intervention at a lower rate than tape measured patients (20.1% vs 27.5%) and fewer patients progressed in this group (7.9% vs 19.2%). The RCT was limited by its open-label design and lack of reporting of important health outcomes. The single prospective comparative study found a significantly lower rate of clinical lymphedema in patients managed with BIS devices but had several limitations, including nonrandomized design, lack of blinding, lack of complete data on a substantial proportion of enrolled patients, and lack of a systematic method for diagnosing lymphedema in the control group. Retrospective studies suggested that postoperative bioimpedance monitoring is feasible but provide limited information about its efficacy. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
Not applicable.
The objective of this evidence review is to evaluate whether the use of bioimpedance spectroscopy devices improves the net health outcome for individuals with known or suspected lymphedema.
Devices using bioimpedance (bioelectrical impedance spectroscopy) are considered investigational for use in the diagnosis, surveillance, or treatment of patients with lymphedema, including use in subclinical secondary lymphedema.
See the Codes table for details.
BlueCard/National Account Issues
State or federal mandates (eg, Federal Employee Program) may dictate that certain U.S. Food and Drug Administration‒approved devices, drugs, or biologics may not be considered investigational, and thus these devices may be assessed only by their medical necessity.
Benefits are determined by the group contract, member benefit booklet, and/or individual subscriber certificate in effect at the time services were rendered. Benefit products or negotiated coverages may have all or some of the services discussed in this medical policy excluded from their coverage.
Lymphedema
Lymphedema is an accumulation of fluid due to disruption of lymphatic drainage. Lymphedema can be caused by congenital or inherited abnormalities in the lymphatic system (primary lymphedema) but is most often caused by acquired damage to the lymphatic system (secondary lymphedema). Breast cancer treatment is one of the most common causes of secondary lymphedema. Both the surgical removal of lymph nodes and radiotherapy are associated with development lymphedema in patients with breast cancer. In a systematic review of 72 studies (N=29612 women), DiSipio et al (2013) reported that approximately 1 in 5 women who survive breast cancer will develop arm lymphedema.1, Risk factors for development of lymphedema that had a strong level of evidence were extensive surgery (ie, axillary-lymph-node dissection, greater number of lymph nodes dissected, mastectomy) and being overweight or obese.
Diagnosis and Staging
A diagnosis of secondary lymphedema is based on history (e.g., cancer treatment, trauma) and physical examination (localized, progressive edema and asymmetric limb measurements) when other causes of edema can be excluded. Imaging, such as MRI, computed tomography, ultrasound, or lymphoscintigraphy, may be used to differentiate lymphedema from other causes of edema in diagnostically challenging cases.
Table 1 lists International Society of Lymphology guidance for staging lymphedema based on "softness" or "firmness" of the limb and the changes with an elevation of the limb.2,
Stage | Description |
Stage 0 (subclinical) | Swelling is not evident and most patients are asymptomatic despite impaired lymphatic transport |
Stage I (mild) | Accumulation of fluid that subsides (usually within 24 hours) with limb elevation; soft edema that may pit, without evidence of dermal fibrosis |
Stage II (moderate) | Does not resolve with limb elevation alone; limb may no longer pit on examination |
Stage III (severe) | Lymphostatic elephantiasis; pitting can be absent; skin has trophic changes |
Management and Treatment
Lymphedema is treated using elevation, compression, and exercise. Conservative therapy may consist of several features depending on the severity of the lymphedema. Patients are educated on the importance of self-care including hygiene practices to prevent infection, maintaining ideal body weight through diet and exercise, and limb elevation. Compression therapy consists of repeatedly applying padding and bandages or compression garments. Manual lymphatic drainage is a light pressure massage performed by trained physical therapists or by patients designed to move fluid from obstructed areas into functioning lymph vessels and lymph nodes. Complete decongestive therapy is a multiphase treatment program involving all of the previously mentioned conservative treatment components at different intensities. Pneumatic compression pumps may also be considered as an adjunct to conservative therapy or as an alternative to self-manual lymphatic drainage in patients who have difficulty performing self-manual lymphatic drainage. In patients with more advanced lymphedema after fat deposition and tissue fibrosis has occurred, palliative surgery using reductive techniques such as liposuction may be performed.
Bioimpedance Spectroscopy
Bioimpedance spectroscopy is based on the theory that the level of opposition to the flow of electric current (impedance) through the body is inversely proportional to the volume of fluid in the tissue. In lymphedema, with the accumulation of excess interstitial fluid, tissue impedance decreases.
Bioimpedance has been proposed as a diagnostic test for this condition. In usual care, lymphedema is recognized clinically or via limb measurements. However, management via bioelectrical impedance spectroscopy has been proposed as a way to implement early treatment of subclinical lymphedema to potentially reduce its severity.
A selection of devices that have been cleared for marketing by the U.S. Food and Drug Administration (FDA) through the 510(k) process to aid in the assessment of lymphedema are summarized in Table 2. Among the FDA-approved bioimpedance devices are SOZO (ImpediMed), MoistureMeterD (Delfin Technologies), and the L-Dex U400 (ImpediMed). The L-Dex U400 was discontinued by its manufacturer in November 2018.
Year | Device | Manufacturer | 510(k) Number | Indication |
2018 | SOZO | ImpediMed (Carlsbad, CA) | K180126 | For adults at risk of lymphedema. Supports the measurement of extracellular fluid volume differences between the limbs and is presented to the clinician on an L-Dex scale as an aid to their clinical assessment of lymphedema. The device is only indicated for patients who will have or who have had lymph nodes, from the axillary and/or pelvic regions, either removed, damaged, or irradiated. |
2015 | MoistureMeterD | Delfin Technologies (Stamford, CT) | K143310 | Supports local assessment of tissue water differences between affected and contralateral non-affected arm tissues to aid in forming a clinical judgment of unilateral lymphedema in women. The device is not intended to make diagnosis or predict arm lymphedema. |
FDA product code: OBH.
This evidence review was created in March 2010 and has been updated regularly with searches of the PubMed database. The most recent literature update was performed through November 27, 2024.
Evidence reviews assess whether a medical test is clinically useful. A useful test provides information to make a clinical management decision that improves the net health outcome. That is, the balance of benefits and harms is better when the test is used to manage the condition than when another test or no test is used to manage the condition.
The first step in assessing a medical test is to formulate the clinical context and purpose of the test. The test must be technically reliable, clinically valid, and clinically useful for that purpose. Evidence reviews assess the evidence on whether a test is clinically valid and clinically useful. Technical reliability is outside the scope of these reviews, and credible information on technical reliability is available from other sources.
Promotion of greater diversity and inclusion in clinical research of historically marginalized groups (e.g., People of Color [African-American, Asian, Black, Latino and Native American]; LGBTQIA (Lesbian, Gay, Bisexual, Transgender, Queer, Intersex, Asexual); Women; and People with Disabilities [Physical and Invisible]) allows policy populations to be more reflective of and findings more applicable to our diverse members. While we also strive to use inclusive language related to these groups in our policies, use of gender-specific nouns (e.g., women, men, sisters, etc.) will continue when reflective of language used in publications describing study populations
The purpose of using BIS in patients who have known, or suspected lymphedema, is to inform a diagnosis of subclinical lymphedema to initiate treatment sooner than with other diagnostic methods.
The following PICO was used to select literature to inform this review.
The relevant population of interest is individuals with known or suspected lymphedema.
The relevant intervention of interest is BIS.
Management via BIS has been proposed as a way to implement early treatment of subclinical lymphedema to potentially reduce its severity.
The relevant comparators of interest are volume displacement and circumferential measurement.
In usual care, lymphedema is recognized clinically or via limb measurements.
Volume is measured using different methods; eg, tape measurements with geometry formulas, perometry, and water displacement.
Objective outcomes of interest include a reduction in limb circumference and/or volume and reduction in the rates of infections (eg, cellulitis, lymphangitis).
Patient-reported outcomes (PROs) of interest include symptoms, quality of life (QOL), and functional measures. A systematic review of PRO instruments and outcomes used to assess QOL in breast cancer patients with lymphedema, Pusic et al (2013) found that most studies included generic PRO instruments or oncology PRO instruments.3, Lymphedema-specific instruments are occasionally used; specifically, the Upper Limb Lymphedema 27 was found to have strong psychometric properties.
There does not appear to be a consensus on minimally clinically important change for either objective outcomes such as changes in arm volume or subjective measures such as changes to patient symptoms or QOL.
The time frame for outcomes varies from months to years after the onset of lymphedema symptoms.
For evaluation of clinical validity of bioimpedance testing, studies that meet the following eligibility criteria were considered:
Reported on the accuracy of the marketed version of the technology (including any algorithms used to calculate scores)
Included a suitable reference standard
Patient/sample clinical characteristics were described
Patient/sample selection criteria were described
For evaluation of clinical utility, comparative controlled prospective trials, with preference for RCTs were considered. In the absence of such trials, comparative observational studies, with preference for prospective studies were considered.
A test must detect the presence or absence of a condition, the risk of developing a condition in the future, or treatment response (beneficial or adverse).
A technology assessment on the diagnosis and treatment of secondary lymphedema, performed for the Agency for Healthcare Research and Quality (AHRQ), was published in 2010.4, The AHRQ assessment identified 8 studies that reported the sensitivity and specificity of tests to diagnose secondary lymphedema. Reviewers noted there is no true criterion standard to grade severity of lymphedema and that limb volume and circumference are used as de facto criterion standards. Two of the 8 selected studies evaluated BIS devices.5,6, Overall, reviewers concluded that, due largely to heterogeneity among studies, the evidence did not permit conclusions on the optimal diagnostic test for detection of secondary lymphedema.
A systematic review by Whitworth et al (2022) evaluated strategies for screening and early intervention in breast cancer patients at risk for lymphedema.7, A total of 12 studies (N=2907) were included. Although 4 RCTs were included, only 1 RCT evaluated BIS (see Ridner et al below). Of the 7 prospective, observational studies identified, 5 evaluated BIS. Although these studies generally point to BIS as a sensitive surveillance technique, this analysis did not synthesize data from the included studies and no quality or bias risk was assessed.
After the AHRQ review, several other studies have evaluated the diagnostic performance of BIS devices for detecting lymphedema. Prospective studies that compared bioelectrical impedance analysis to a reference standard are described next.
A study by Barrio et al (2015) enrolled 223 women with newly diagnosed breast cancer and a plan for unilateral axillary surgery.8, Thirty-seven patients were excluded due to ineligibility or withdrawal, leaving a sample size of 186. Prior to surgery, participants received baseline volumetric measurements with a bioimpedance device (L-Dex) and volume displacement (the reference standard). Patients then had follow-up volumetric measurements every 3 to 6 months for 3 years. At the last follow-up (median, 18.2 months), 152 (82%) patients had no lymphedema, 21 (11%) had an abnormal L-Dex, and no lymphedema by volume displacement, 4 (2%) had an abnormal L-Dex and lymphedema by volume displacement, and 9 (5%) had lymphedema without prior L-Dex abnormality. In an analysis including only patients with at least 6 months of follow-up, L-Dex had a sensitivity of 31% (4/13) and a specificity of 88% (129/147) for predicting subsequent lymphedema development. Also, the correlation between changes in volume displacement and changes in L-Dex results were in the low-to-moderate range at 3 months (r=0.31) and 6 months (r=0.21). However, at the time of lymphedema diagnosis, the L-Dex ratio was abnormal in 12 of 13 patients (diagnostic sensitivity, 92%).
Blaney et al (2015) reported on a prospective study with 126 women with stage I, II, or III unilateral breast cancer.9, A total of 115 women underwent baseline assessment with an L-Dex and circumferential measurement. The circumferential measurement was used as the reference standard, although the authors noted the test is an imperfect criterion standard. Postsurgical follow-up assessments were planned every 3 months for a year. The number of women completing these assessments was 109 (95%) at 3 months, 89 (77%) at 6 months, 79 (69%) at 9 months, and 71 (62%) at 12 months. Over 12 months, 31 participants were identified as having lymphedema by at least 1 of the assessment methods. Twenty-eight (90%) of 31 were identified by circumferential measurement and 11 (35%) by BIS. There was no statistically significant correlation between bioimpedance analysis and circumferential measurement.
A test is clinically useful if the use of the results informs management decisions that improve the net health outcome of care. The net health outcome can be improved if individuals receive correct therapy, or more effective therapy, or avoid unnecessary therapy, or avoid unnecessary testing.
The ideal study design is an RCT comparing health outcomes in individuals managed with and without the use of bioimpedance devices.
One multicenter, international, RCT conducted by Ridner et al (2019 and 2022) [PREVENT RCT] compared bioimpedance to volume measurements calculated from arm circumference using a tape measure (Table 3).10,11, The primary aim of the study was to determine if subclinical detection of extracellular fluid accumulation via BIS and subsequent early intervention reduces the rate of progression to clinical lymphedema relative to the rates seen using standard tape measurements. Patients requiring early intervention were prescribed a compression sleeve and gauntlet for 4 weeks and then re-evaluated. Predetermined thresholds were used to trigger early intervention. The implementation threshold for patients in the bioimpedance group was initially a change that was ≥10 L-Dex units (3 standard deviations) higher than the presurgical baseline measure, but the protocol was changed in 2016 to include all patients with ≥6 L-Dex units. Patients in the tape measure group triggered when they had a volume change in the at-risk arm that was between >5 and <10% above the presurgical baselines. Progression to clinical lymphedema was defined as a 10% or greater increase in tape measure volume from baseline in the at-risk arm.
Results of the interim analysis and final analysis are summarized in Table 4.10,11, At interim analysis , 109 of 508 (21.9%) patients received early intervention due to reaching the pre-determined threshold. Patients randomized to bioimpedance had a lower rate of trigger and longer times to trigger. A total of 12 triggering patients progressed to clinical lymphedema (10 in the TM group [14.7%] and 2 in the BIS group [4.9%]). The difference between groups was not statistically significant (p=.130) and did not meet stopping criteria specified in the study protocol. At final analysis (median of 32.9 months follow-up), BIS triggered an intervention at a lower rate than TM patients (20.1% vs 27.5%; p=.011); however, fewer patients in the BIS group progressed compared with tape measure (7.9% vs 19.2%; relative risk, 0.41; 95% CI, 2.8-4.5; p=.001).
This study had several limitations (see Tables 5 and 6), including an open-label design, which may have introduced bias in outcome assessment, treatments, or the decision to trigger an intervention. Important health outcomes such as patient-reported symptoms, QOL, and function were not assessed. Additionally, 39 patients who progressed prior to an intervention being triggered were excluded from the analysis.
Shah et al. (2024) conducted a secondary analysis on data from the PREVENT RCT to investigate the onset and progression of subclinical breast cancer-related lymphedema (sBCRL) and clinical breast cancer-related lymphedema (cBCRL).12,The aim was to provide guidance on the optimal screening frequency and duration for BCRL. A cohort of 919 women at risk of developing cBCRL were regularly screened using either bioimpedance or tape measure (TM) for up to 36 months following their breast cancer treatment. Women at risk of cBCRL (N=919) were regularly screened for up to 36 months post breast cancer treatment using either bioimpedance or TM. In total, 209 patients (23%) developed sBCRL (bioimpedance: n=89, TM: n=120) and were eligible for intervention. Subsequently, 30 patients progressed to cBCRL post-intervention (BIS: 7, TM: 23). More than half of the patients exhibited measurements consistent with sBCRL within 9 months of breast cancer treatment. Initial detections of sBCRL persisted, regardless of the screening method used, with rates remaining stable in the second and third years (p>0.24) post-surgery. Furthermore, 39 patients progressed to cBCRL without previously developing sBCRL or receiving intervention over the 3-year period. The timing of sBCRL detection highlights that patients remain at risk years after treatment and may continue to progress to cBCRL long after surgery. Early detection of sBCRL facilitates timely intervention, thereby reducing the likelihood of progression to cBCRL. Consequently, patients should be diligently monitored for a minimum of 3 years following the completion of cancer treatment, with particular emphasis on focused and targeted monitoring during the initial 9-month period.
Study; Trial | Country | Sites | Dates | Participants | Interventions | |
Active | Comparator | |||||
Ridner et al (2019 and 2022)10,11,PREVENT-NCT02167659 | U.S. and Australia | 13 | 2014-2018 | Presurgical: Women >18 years of age with histologically confirmed, newly diagnosed, breast cancer (invasive or DCIS) with planned surgery. Postsurgical: stage I–III invasive breast cancer or DCIS who received ≥1 of the following: mastectomy, axillary treatment, regional node irradiation, or taxane-based chemotherapy | BIS: N=263 at interim; 482 at final | Tape measure: N=245 at interim; 481 at final |
BIS: bioimpedance spectroscopy; DCIS: ductal carcinoma in situ; NCT: national clinical trial; PREVENT: Bioimpedance Spectroscopy Versus Tape Measure in Prevention of Lymphedema; RCT: randomized controlled trial.
Study | Intervention triggered | Median (IQR) months to intervention triggered | Progression to clinical lymphedema | Median (range) months to progression to clinical lymphedema |
Ridner et al (2019)10, | ||||
BIS | 41/259 (15.8%) | 2.8 (0.6–5.6) | 2/41 (4.9%) | 6.0 (1.4, 16.9) |
Tape measure | 68/239 (28.5%) | 4.0 (1.0–11.2) | 10/68 (14.7%) | 6.0 (0.8, 16.9)) |
p-value | .001 | .002 | .130 | .389 |
Ridner et al (2022)11, | ||||
BIS | 89/442 (20.1%) | 9.7 (3.6-18.2) | 7/89 (7.9%) | 4.9 (0.7-15.2) |
Tape measure | 120/437 (27.5%) | 3.9 (1.0-11.6) | 23/120 (19.2%) | 10.7 (1.4-31.9) |
p-value | .011 | .001 | .016 | .100 |
BIS: bioimpedance spectroscopy; IQR: interquartile range; RCT: randomized controlled trial.
Study | Populationa | Interventionb | Comparatorc | Outcomesd | Duration of Follow-Upe |
Ridner et al (2019 and 2022)10,11, | 1. Patient-reported outcomes not assessed |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment. a Population key: 1. Intended use population unclear; 2. Clinical context is unclear; 3. Study population is unclear; 4. Study population not representative of intended use. b Intervention key: 1. Classification thresholds not defined; 2. Version used unclear; 3. Not intervention of interest. c Comparator key: 1. Classification thresholds not defined; 2. Not compared to credible reference standard; 3. Not compared to other tests in use for same purpose. d Outcomes key: 1. Study does not directly assess a key health outcome; 2. Evidence chain or decision model not explicated; 3. Key clinical validity outcomes not reported (sensitivity, specificity, and predictive values); 4. Reclassification of diagnostic or risk categories not reported; 5. Adverse events of the test not described (excluding minor discomforts and inconvenience of venipuncture or noninvasive tests). e Follow-Up key: 1. Follow-up duration not sufficient with respect to natural history of disease (true-positives, true-negatives, false-positives, false-negatives cannot be determined).
Study | Selectiona | Blindingb | Delivery of Testc | Selective Reportingd | Data Completenesse | Statisticalf |
Ridner et al (2019 and 2022)10,11, | 1. Open-label | 2. 10 patients who progressed prior to triggered intervention were excluded from interim and 39 from final analysis |
The study limitations stated in this table are those notable in the current review; this is not a comprehensive gaps assessment. a Selection key: 1. Selection not described; 2. Selection not random or consecutive (ie, convenience). b Blinding key: 1. Not blinded to results of reference or other comparator tests. c Test Delivery key: 1. Timing of delivery of index or reference test not described; 2. Timing of index and comparator tests not same; 3. Procedure for interpreting tests not described; 4. Expertise of evaluators not described. d Selective Reporting key: 1. Not registered; 2. Evidence of selective reporting; 3. Evidence of selective publication. e Data Completeness key: 1. Inadequate description of indeterminate and missing samples; 2. High number of samples excluded; 3. High loss to follow-up or missing data. f Statistical key: 1. Confidence intervals and/or p values not reported; 2. Comparison with other tests not reported.
One prospective observational study compared clinical lymphedema rates in patients managed with and without bioimpedance analysis. This study, by Soran et al (2014), involved prospective detection of subclinical lymphedema in 186 women with breast cancer managed with L-Dex or tape measurement of limb circumference.13, Measurements were obtained at baseline and 3- to 6-month intervals for 5 years. Subclinical lymphedema was defined as an L-Dex value outside the normal range, or that increased at least 10 units from baseline. Patients diagnosed with subclinical lymphedema were treated with, eg, short-term physical therapy, compression garments, and received education on exercise and limb elevation. A total of 180 women were included in the analysis. Seventy-two women had both preoperative and postoperative bioimpedance and tape measurements (preoperative group). Forty-four women had preoperative bioimpedance and tape measurements but only had tape measurements postoperatively (control group). The remaining 64 women had postoperative bioimpedance and tape measurements, but no preoperative measurements (no preoperative group). The authors compared the demographic and clinical characteristics of the preoperative and control groups and the preoperative and postoperative groups; they did not identify any statistically significant differences.
In the preoperative group, 28 (36%) of 72 women were diagnosed with subclinical lymphedema and referred for treatment; 2 women progressed to clinical lymphedema. In the control group, 16 women (36%) developed clinical lymphedema during follow-up. Limitations of the study included a lack of an alternative method for detecting subclinical lymphedema in women in the control group so that they could receive treatment early; a lack of randomization to a treatment group; and incomplete data on pre- and postoperative measures of lymphedema except in a subset of the total population.
Multiple uncontrolled observational studies have reported rates of lymphedema identified through surveillance with bioimpedance in women at high-risk following breast cancer treatment.14,15,16,17,18,19,20,21,22,23, Because these studies did not include a comparison group of women who received usual care or alternative methods of screening, they do not provide evidence to draw conclusions about the clinical utility of bioimpedance.
A 2010 AHRQ technology assessment identified few studies on BIS analysis for diagnosing lymphedema. A 2022 systematic review identified the PREVENT trial as the primary evidence evaluating BIS for lympedema surveillance. Results from the PREVENT RCT comparing BIS with standard tape measure following treatment for breast cancer have been published. At a median follow-up of 32.9 months, BIS patients triggered intervention at a lower rate than tape measure patients (20.1% vs 27.5%) and fewer patients progressed in this group (7.9% vs 19.2%). Additional studies to confirm these findings are needed, especially RCTs and trials that include an alternative method for early or subclinical lymphedema detection.
[ ] MedicallyNecessary | [X] Investigational |
For individuals who have known or suspected lymphedema who receive bioimpedance spectroscopy (BIS), the evidence includes systematic reviews, 1 randomized controlled trial (RCT), 1 prospective comparative observational study, and multiple uncontrolled observational studies. Relevant outcomes are test validity, symptoms, and quality of life. Diagnostic accuracy studies have found a poor correlation between bioimpedance analysis and the reference standard (volume displacement or circumferential measurement). Results from the PREVENT RCT comparing bioimpedance with standard tape measure following treatment for breast cancer have been published. At a median follow-up of 32.9 months, BIS patients triggered intervention at a lower rate than tape measured patients (20.1% vs 27.5%) and fewer patients progressed in this group (7.9% vs 19.2%). The RCT was limited by its open-label design and lack of reporting of important health outcomes. The single prospective comparative study found a significantly lower rate of clinical lymphedema in patients managed with BIS devices but had several limitations, including nonrandomized design, lack of blinding, lack of complete data on a substantial proportion of enrolled patients, and lack of a systematic method for diagnosing lymphedema in the control group. Retrospective studies suggested that postoperative bioimpedance monitoring is feasible but provide limited information about its efficacy. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
The purpose of the remaining sections in Supplemental Information is to provide reference material regarding existing practice guidelines and position statements, U.S. Preventive Services Task Force Recommendations and Medicare National Coverage Decisions and registered, ongoing clinical trials. Inclusion in the Supplemental Information does not imply endorsement and information may not necessarily be used in formulating the evidence review conclusions.
While the various physician specialty societies and academic medical centers may collaborate and make recommendations during this process, through the provision of appropriate reviewers, input received does not represent an endorsement or position statement by the physician specialty societies or academic medical centers, unless otherwise noted.
In response to requests, input was received from 2 specialty societies and 2 academic medical centers while this policy was under review in 2011. Three of 4 reviewers agreed that bioimpedance devices are considered investigational for diagnosis, surveillance, and treatment of patients with lymphedema. The fourth reviewer, from an academic medical center, considered the use of the technology a reasonable alternative, especially in situations in which minor lymphedema can have a large impact on a patient. One specialty society supported further research into the effectiveness of this technology and recommended reimbursement in the context of relevant clinical trials.
Guidelines or position statements will be considered for inclusion in ‘Supplemental Information’ if they were issued by, or jointly by, a US professional society, an international society with US representation, or National Institute for Health and Care Excellence (NICE). Priority will be given to guidelines that are informed by a systematic review, include strength of evidence ratings, and include a description of management of conflict of interest.
National Comprehensive Cancer Network (NCCN) Clinical Practice Guidelines on Survivorship (v.1.202 4) recommends that survivors at risk for lymphedema should be regularly screened for lymphedema by symptom assessment, clinical exam, and, if available, bioimpedance spectroscopy.24,
NCCN Clinical Practice Guidelines on Breast Cancer (v.6.202 4 ) recommend education, monitoring, and referral for lymphedema management as needed. For further information they refer the reader to the Survivorship Guidelines.25,
Not applicable.
There is no national coverage determination. In the absence of a national coverage determination, coverage decisions are left to the discretion of local Medicare carriers.
Some currently ongoing and unpublished trials that might influence this review are listed in Table 7.
NCT No. | Trial Name | Planned Enrollment | Completion Date |
Ongoing | |||
NCT01521741 | Prospective Screening for Breast Cancer-related Lymphedema: Analysis of Objective Measurements, Symptoms, Functionality, and Quality of Life Questionnaires to Evaluate Lymphedema in Patients Following Treatment for Breast Cancer. | 10000 | Dec 2026 |
NCT03292198a | Treatment Indications for Breast Cancer-related Subclinical Lymphedema Identified Through a Bioimpedance Surveillance Model | 267 | Dec 2025 |
NCT03978754 | Assessment of Breast Cancer-Related Arm Lymphedema—Comparison of Traditional Measurement Methods and Indocyanine Green (ICG) Lymphography | 1600 | Jan 2022 (status unknown ) |
NCT02743858 | A Prospective Surveillance Program for Assessment and Treatment of Breast Cancer-Related Lymphedema After Axillary Lymph Node Dissection | 1250 | Apr 2025 (Recruiting as of Oct 2024 ) |
a Denotes industry-sponsored or cosponsored trial; BIS: bioimpedance spectroscopy; NCT: national clinical trial.
Codes | Number | Description |
---|---|---|
CPT | 93702 | Bioimpedance spectroscopy (BIS), extracellular fluid analysis for lymphedema assessment(s) |
ICD-10-CM | Investigational for all diagnoses | |
I97.2 | Postmastectomy lymphedema syndrome | |
Z90.10-Z90.13 | Acquired absence of breast and nipple code range | |
ICD-10-PCS | ICD-10-PCS codes are only used for inpatient services. There is no specific ICD-10-PCS code for this testing. | |
Type of Service | Medical | |
Place of Service | Outpatient |
N/A
Date | Action | Description |
---|---|---|
02/10/2025 | Annual Review | Policy updated with literature review through November 27, 2024; references added. Policy statement unchanged. |
02/09/2024 | Annual Review | Policy updated with literature review through November 27, 2023; references added. Policy statement unchanged. |
02/07/2023 | Annual Review | Policy updated with literature review through October 17, 2022; no references added. Minor editorial refinements to policy statement; intent unchanged. |
02/07/2022 | Annual Review | Policy updated with literature review through October 18, 2021; no references added; NCCN guidelines updated. Policy statement unchanged. |
02/01/2021 | Annual Review | Policy updated with literature review through October 19, 2020; no references added. Policy statement unchanged. |
02/24/2020 | Policy Reviewed | Policy updated with literature review through November 14, 2019; references added. Policy statement unchanged. |
02/20/2019 | Annual Review | Update |
01/11/2018 | ||
06/23/2017 | ||
01/14/2016 |